<5. Functions of CQ1 Software> 5-31
IM 80J01A01-01E
(5) Objective Lens
Select objective lens here. Selected lens is highlighted.
(6) Time-lapse Setting
Switch the time-lapse acquisition mode (normal mode/ fast mode
(option)).
If you didn’t buy this option, this menu isn’t shown and the time-lapse
acquisition mode is always normal mode.
󳏀 Normal
In this mode, sequence of imaging in each well, field, channel and
Z slice has a priority. These sequences of imaging are performed
at interval and duration set at Time-lapse Condition described
below.
󳏁 Fast (option)
In this mode, time-lapse imaging has a priority. Time-lapse
imaging of speed and time length set at Fast Time-lapse
Condition is performed in each well and field. Numbers of channel
and Z slice are fixed to 1. When switching to this mode, the
selected channel remains and the others are removed
automatically. Also, Chromatic Abberation becomes invalid
automatically.
This mode is not correspond to on-the-fly analysis. To analyze
image data acquired by this mode, it is necessary to acquire image
data by CQ mode then analyze by HDD mode.
󳏀
󳏁
Note
Dont select Fast mode if setting of folder structure of saving
data is By Time Point (Refer to 5.11 User Setting).
In case to use Fast mode, select Main Menu -> “Option ->
User Setting and select setting other than By Time Point.