1-87
IM 04L51B01-01EN
Configuring the GX/GP and Viewing the Settings
1
2
3
4
5
App
Index
Reset
To reset the TLOG computed value at each interval, select On.
The figure below illustrates the reset operation of sum computation (TLOG.SUM).
Example: Result of the TLOG.SUM computation
When reset is On, the sum value is computed over each interval. When set to Off, the sum
value from computation start is computed.
1 2 3
ResetReset Reset
Reset: On
Timer timeout
Reset: Off
Rolling average
Setup Item Selectable Range or Options Default Value
On/Off Off, On Off
Interval
1
1s, 2s, 3s, 4s, 5s, 6s, 10s, 12s, 15s, 20s, 30s,
1min, 2min, 3min, 4min, 5min, 6,in, 10min,
12min, 15min, 20min, 30min, 1h
10s
Number of samples
1
1 to 1500 1
1 Appears when rolling average is On.
On/Off
Set this to On to perform rolling average.
Interval
Set the sampling interval for performing rolling average. The sampling interval takes on a
value that is an integer multiple of the scan interval. For example, if the sampling interval is
set to 5 s when the scan interval is 2 s, the actual sampling interval is 6 s.
Number of samples
Set the number of samples for performing rolling average.
The rolling average time is equal to the sampling interval × the number of samples.
Note
If the number of data points to be averaged has not reached the specified number of samples
immediately after computation is started, the average of the available data is computed.
Computation error data is excluded from the rolling average computation.
If the computed data exceeds the upper or lower limit, the data is handled according to the
Value on Overflow setting.
If it is set to Error or Skip, the computed data is not used in the rolling average computation.
If set to Limit, the computed data is clipped at the upper or lower limit to compute the rolling
average.
The upper and lower limits are “±100000000” ignoring the decimal point. The decimal place is
the same as that of the span lower limit.
1.8ConfiguringMathChannels(/MToption)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA GX20 and is the answer not in the manual?

YOKOGAWA GX20 Specifications

General IconGeneral
BrandYOKOGAWA
ModelGX20
CategoryRecording Equipment
LanguageEnglish

Related product manuals