EasyManuals Logo

YOKOGAWA GX20 User Manual

YOKOGAWA GX20
707 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #629 background imageLoading...
Page #629 background image
17
All Rights Reserved. Copyright © 2012, Yokogawa Electric Corporation
GS 04L51B01-01EN
June. 14, 2017-00
Limitations
Systemlimitations
Maximumnumberofchannels
System Measurment
mode
Maximum
number
ofmodule
connections
Maximumnumberofchannels
Input/output Communication Math Report
GX10 Normal 10 100 50 50 50
High speed 3 12 50 50 50
Dualinterval 10 50 25 25 25
GX20-1 Normal 10 100 300 100 60
High speed 10 40 300 100 60
Dualinterval 10 50 150 50 30
GX20-2 Normal 45 500 500 200 60
High speed 10 40 500 200 60
Dualinterval 45 250 250 100 30
LimittothenumberofexpandableI/OunitsorGMsubunits
•Uptosixunitscanbeconnected.
•ConnectionisnotpossibleifthemeasurementmodeissettoHighspeed.
LimittothenumberofGXmainunitmodules
Module GX10 GX20-1 GX20-2
WhenGX90XA-04-H0andGX90YAare
included
No limit 9 9
WhenGX90UTisincluded No limit 8 8
LimitonModules
•Upto10modulesconsistingofGX90YD,GX90WD,andGX90UTcanbeinstalledintothesystem.
•OneGX90WDmodulecanbeinstalledinaGX.OnemodulecanbeinstalledinaGX60(expandableI/O)andeach
GM sub unit.
•OneGX90YAmodulecanbeinstalledinaGX10.TwomodulescanbeinstalledineachoftheGX20,GX60
(expandableI/O)andGMsubunit.
•Upto10GX90YAmodulescanbeinstalledinaGX10/GX20-1systemandupto12inaGX20-2system.
•IfthemeasurementmodeisHighspeed,aGX90XDorGX90WDmodulecanbeinstalledinthesystem.
•IfthemeasurementmodeisHighspeed,onlyGX90XA-04-H0(high-speedAI),GX90XD(DI),andGX90WD(DIO)
aredetected.DIandDIOarexedtoremotemode.Measurementandrecordingarenotpossible.
•IfthemeasurementmodeisDualinterval,GX90UTisnotdetected.
LimittotheNumberofMeasurementChannels
Thefollowinglimitationappliestothenumberofchannelsthatcanmeasureatscanintervalsshorterthan100ms.
IfthemeasurementmodeisDualinterval,thelimitationappliesatthescangrouplevel.However,therearescan
intervals you cannot set.
● MeasurementmodeisHighspeed
GX10
Channel Scaninterval
1ms 2ms 5ms 10ms 20ms 50ms
Input/output 1 2 5 10 12 12
math 1 2 5 10 25
communication 1 2 5 10 25
Report 1 2 5 10 25