11-2 Nightingale PPM3 User's Guide April 2019
Working With Menus
11.1. Parameter Menus
11.1.1. ECG and Heart Rate
Figure 31. Setup HR Menu
Setting
Description
Lower Limit
HR/PR alarm lower limit
Upper Limit
HR/PR alarm upper limit
Auto
Auto-adjust lower limit to 80% of current value and upper limit to
125% of current value.
Alarms On
Enables (Yes) or disables (No) alarms
Print on Alarm
Print a strip chart summary upon alarm condition
ECG Lead
Select lead used for HR processing
Size
Changes display amplitude of ECG waveform

Table of Contents