April 2019 Nightingale PPM3 User's Guide 19-1
PPM3 Monitor Settings
19. PPM3 Monitor Settings
The following tables show the factory default settings (Patient Data, Parameter,
Waveform, and Device) for the Nightingale PPM3 monitor. Note the differences
in certain parameter settings for Adult and Pediatric patient types. You may
adjust the parameter and waveform settings for your particular application, and
then save this configuration as a department default for all new patients. These
department default settings can be adjusted as needed for specific patients, as
shown in the Working With Menus chapter. System Settings are specific to a
particular PPM3 device.
Patient Data Settings
Setting name
Default
value
Possible values
Patient Last Name
<BLANK>
any text string (39 characters maximum)
Patient Middle Initial
<BLANK>
any text string (1 characters maximum)
Patient First Name
<BLANK>
any text string (39 characters maximum)
Patient ID
<BLANK>
any text string (39 characters maximum) but
may not include the following characters (since
they are not legal Windows file names, and the
Patient ID is used to create folder and file
names):
\ / : * ? " < > |
Patient Type
Adult
Adult, Pediatric
Patient Sex
<BLANK>
Unknown (BLANK), Male, Female
Patient Date Of Birth Year
<BLANK>
(Current Year 150) up to Current Year
Patient Date Of Birth Month
<BLANK>
1 - 12
Patient Date Of Birth Day
<BLANK>
1 maximum number of days in specified
month and year
Patient Height
<BLANK>
0 157 inches
0 400 cm
Patient Weight
<BLANK>
0 1100 lbs
0 500 kg

Table of Contents