19-2 Nightingale PPM3 User's Guide April 2019
PPM3 Monitor Settings
Parameter Settings
Setting name
Default value
Adult / Pediatric
Possible values
ETCO
2
Enabled
No
---
Yes, No
IBP Channels
None
---
None, 1, 2
Numbers Only
No
---
Yes, No
HR Lower Alarm Limit
50 bpm
---
15 to 299 bpm, Off
HR Upper Alarm Limit
120 bpm
150 bpm
16 to 300 bpm, Off
HR Alarms On
Yes
---
Yes, No
HR Print on Alarm
No
---
Yes, No
ECG Lead
II
---
I, II, III, V, aVL, aVR, aVF
(ECG Lead) Size
10 mm/mV
---
2.5 mm/mV, 5 mm/mV, 10 mm/mV, 15 mm/mV
HR Source
ECG
---
ECG, SpO2, ART
Pulse Tone Source
Off
---
ECG, SpO2, Off
ECG Filter
Monitor
(0.67 40
Hz)
---
Monitor (0.67 40 Hz),
Diagnostic (0.05 40 Hz)
RR Lower Alarm Limit
5 bpm
10 bpm
5 to 149 bpm, Off
RR Upper Alarm Limit
20 bpm
45 bpm
6 to 150 bpm, Off
RR Alarms On
Yes
---
Yes, No
RR Print on Alarm
No
---
Yes, No
ETCO
2
Lower Alarm Limit
35 mmHg
---
0 to 145 mmHg, Off
ETCO
2
Upper Alarm Limit
45 mmHg
---
5 to 150 mmHg, Off
ETCO
2
Alarms On
Yes
---
Yes, No
ETCO
2
Print on Alarm
No
---
Yes, No
FICO
2
Upper Alarm Limit
2 mmHg
---
0 to 98 mmHg, Off
FICO
2
Alarms On
Yes
---
Yes, No
FICO
2
Print on Alarm
No
---
Yes, No
RRc Lower Alarm Limit
5 bpm
10 bpm
5 to 149 bpm, Off
RRc Upper Alarm Limit
20 bpm
45 bpm
6 to 150 bpm, Off
RRc Alarms On
Yes
---
Yes, No
RRc Print on Alarm
No
---
Yes, No
(CO2 waveform) Size
0 to 40
mmHg
---
0 to 20 mmHg, 0 to 40 mmHg, 0 to 60 mmHg, 0 to 80
mmHg
Apnea Time
20 seconds
---
10, 15, 20, 25, 30 seconds
O
2
Compensation (Masimo)
0 to 30
---
0 to 30, 31 to 70, 71 to 100

Table of Contents