April 2019 Nightingale PPM3 User's Guide 10-3
10.2. Temperature Monitoring Messages
Note – Factory default Temperature alarm and limit settings may be found in the Menu
Settings table in the PPM3 Monitor Settings chapter.
The patient's
temperature has fallen
below the current lower
alarm limit.
Check the patient and provide any
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
The patient's
temperature has risen
above the current upper
alarm limit.
Check the patient and provide any
necessary clinical care.
Change the alarm limit if it is no longer
clinically appropriate.
Temperature probe
disconnected
Check to make sure the temperature
probe is connected to the temperature
cable.
Check to make sure the temperature
cable is connected to the PPM3.
The patient's
temperature has risen
above the maximum
value the monitor can
accurately detect.
There is a problem with
the connections or with
the hardware.
Check the patient and provide any
necessary clinical care.
Check the temperature cable
connections.
Turn the monitor off, then on.
If message persists, contact Zoe
technical support.
Monitor has detected a
hardware problem.
Turn the monitor off, then on.
If message persists, contact Zoe
technical support.