April 2019 Nightingale PPM3 User's Guide 11-9
Working With Menus
11.1.5. Oridion CO
2
/ Capnography
Figure 36. Setup CO
2
Menu With Oridion Capnography
Setting
Description
Lower Limit
ETCO
2
, RRc, & IPI alarm lower limits
Upper Limit
ETCO
2
, RRc, & FICO
2
alarm upper limits
Auto
Auto-adjust lower limits to 80% of current values and upper limits to
125% of current values.
Alarms On
Enables (Yes) or disables (No) alarms
Print on Alarm
Print a strip chart summary upon alarm conditions
Size
Changes display amplitude of CO
2
waveform
Apnea Time
Time since last breath before RRc alarm annunciates
IPI Enabled
Enables (Yes) or disables (No) Integrated Pulmonary Index
Age Range
Required age range for IPI use with pediatric patient