April 2019 Nightingale PPM3 User's Guide 11-13
Working With Menus
11.1.8. Temperature
Figure 39. Setup TEMP Menu
Setting
Description
Lower Limit
TEMP alarm lower limit
Upper Limit
TEMP alarm upper limit
Auto
Auto-adjust lower limit to 95% of current value and upper limit to
105% of current value.
Alarms On
Enables (Yes) or disables (No) alarms
Print on Alarm
Print a strip chart summary upon alarm conditions
Note When you change the TEMP Units setting (°F or °C) from the Setup System
menu, recheck the TEMP alarm limit settings, as some round-off error may occur when
the limit settings are switched to the new units of measure.