April 2019 Nightingale PPM3 User's Guide 19-3
PPM3 Monitor Settings
Parameter Settings
Setting name
Default value
Adult / Pediatric
Possible values
N
2
O Compensation (Masimo)
0 to 30
---
0 to 30, 31 to 70
IPI Enabled (Oridion)
Yes
---
Yes, No
IPI Lower Limit
4
---
1 to 9, Off
IPI Alarms On
No
---
Yes, No
IPI Print on Alarm
No
---
Yes, No
IPI Pediatric Age Range
N/A
<BLANK>
1 to 3 years, 3 to 6 years, 6 to 12 years
SpO
2
Lower Alarm Limit
90%
---
50 to 99%, Off
SpO
2
Upper Alarm Limit
100%
---
51 to 100%, Off
SpO
2
Alarms On
Yes
---
Yes, No
SpO
2
Print on Alarm
No
---
Yes, No
P1s Lower Alarm Limit
90 mmHg
---
-10 to 249 mmHg, Off
P1s Upper Alarm Limit
180 mmHg
145 mmHg
-9 to 250 mmHg, Off
P1s Alarms On
No
---
Yes, No
P1s Print on Alarm
No
---
Yes, No
P1m Lower Alarm Limit
75 mmHg
---
-10 to 229 mmHg, Off
P1m Upper Alarm Limit
110 mmHg
---
-9 to 230 mmHg, Off
P1m Alarms On
No
---
Yes, No
P1m Print on Alarm
No
---
Yes, No
P1d Lower Alarm Limit
55 mmHg
---
-10 to 209 mmHg, Off
P1d Upper Alarm Limit
100 mmHg
---
-9 to 210 mmHg, Off
P1d Alarms On
No
---
Yes, No
P1d Print on Alarm
No
---
Yes, No
P1 Label
P1
---
P1, ART, PA, CVP
(P1 waveform) Size
0 to 200
mmHg
---
-10 to 20 mmHg, -10 to 60 mmHg, 0 to 150 mmHg,
0 to 200 mmHg, 0 to 250 mmHg, 0 to 300 mmHg
P1 Format
SYS/DIA
---
SYS/DIA, MEAN
P2s Lower Alarm Limit
90 mmHg
---
-10 to 249 mmHg, Off
P2s Upper Alarm Limit
180 mmHg
145 mmHg
-9 to 250 mmHg, Off
P2s Alarms On
No
---
Yes, No
P2s Print on Alarm
No
---
Yes, No
P2m Lower Alarm Limit
75 mmHg
---
-10 to 229 mmHg, Off
P2m Upper Alarm Limit
110 mmHg
---
-9 to 230 mmHg, Off