19-4 Nightingale PPM3 User's Guide April 2019
PPM3 Monitor Settings
Parameter Settings
Setting name
Default value
Adult / Pediatric
Possible values
P2m Alarms On
No
---
Yes, No
P2m Print on Alarm
No
---
Yes, No
P2d Lower Alarm Limit
55 mmHg
---
-10 to 209 mmHg, Off
P2d Upper Alarm Limit
100 mmHg
---
-9 to 210 mmHg, Off
P2d Alarms On
No
---
Yes, No
P2d Print on Alarm
No
---
Yes, No
P2 Label
P2
---
P2, ART, PA, CVP
(P2 waveform) Size
0 to 200
mmHg
---
-10 to 20 mmHg, -10 to 60 mmHg, 0 to 150 mmHg,
0 to 200 mmHg, 0 to 250 mmHg, 0 to 300 mmHg
P2 Format
SYS/DIA
---
SYS/DIA, MEAN
TEMP Lower Alarm Limit (Note:
Depends On User Setting For
TEMP Units)
36.0 C
97.0 F
---
32.0 to 41.9 degrees C, Off
89.6 to 107.5 degrees F, Off
TEMP Upper Alarm Limit (Note:
Depends On User Setting For
TEMP Units)
39.0 C
102.0 F
---
32.1 to 42.0 degrees C, Off
89.7 to 107.6 degrees F, Off
TEMP Alarms On
No
---
Yes, No
TEMP Print on Alarm
No
---
Yes, No
NBPs Lower Alarm Limit
90 mmHg
---
30 to 249 mmHg, Off
NBPs Upper Alarm Limit
180 mmHg
145 mmHg
31 to 250 mmHg, Off
NBPs Alarms On
No
---
Yes, No
NBPs Print on Alarm
No
---
Yes, No
NBPm Lower Alarm Limit
75 mmHg
---
20 to 229 mmHg, Off
NBPm Upper Alarm Limit
110 mmHg
---
21 to 230 mmHg, Off
NBPm Alarms On
No
---
Yes, No
NBPm Print on Alarm
No
---
Yes, No
NBPd Lower Alarm Limit
55 mmHg
---
10 to 209 mmHg, Off
NBPd Upper Alarm Limit
100 mmHg
---
11 to 210 mmHg, Off
NBPd Alarms On
No
---
Yes, No
NBPd Print on Alarm
No
---
Yes, No
NBP Interval
Off
---
Off, 1, 2, 3, 5, 10, 15, 30, 60, 120 minutes
NBP Initial Inflation Pressure
160
mmHg
120 mmHg
100 to 270 mmHg in steps of 10 mmHg
Trend Interval
15 min
---
1 min, 5 min, 15 min, 1 h, 4 h