Indications for Use
The Zoe Medical Nightingale Monitoring System is indicated for use in adult &
pediatric patient populations.
The Zoe Medical Nightingale Monitoring System facilitates the monitoring of:
- ECG
- Impedance respiration
- Non-Invasive blood pressure
- Invasive blood pressure
- Body temperature
- Functional arterial oxygen saturation (SpO
2
)
- End-tidal & inspired CO
2
The Zoe Medical Nightingale Monitoring System is a prescription device
intended to be used by healthcare professionals in all areas of a healthcare facility.
WARNING – The Nightingale PPM3 is not intended for use in the
following cases:
• Neonatal patients
• Apnea monitoring
• In an MRI environment
• Applications requiring automated arrhythmia detection
• Applications requiring diagnostic-quality ECG
WARNING – DEFIBRILLATION. To avoid the possibility of serious
injury or death during patient defibrillation, do not come into contact
with patient monitor or patient cables. Additionally, proper
placement of defibrillator paddles in relation to the ECG electrodes is
required to minimize harm to the patient. Only use accessories
approved by Zoe Medical for use with the Nightingale PPM3.
WARNING – PACEMAKER PATIENTS. Rate meters may continue to
count the pacemaker rate during occurrences of cardiac arrest or
some arrhythmias. Do not rely entirely upon heart rate meter alarms.
Keep pacemaker patients under close surveillance. See the
Technical Data chapter for disclosure of the pacemaker pulse
rejection capability of this instrument.