April 2019 Nightingale PPM3 User's Guide 11-31
Working With Menus
The settings that can be changed via the Setup Alarms menu include:
Setting
Description
Alarm Silence Time
Time that alarm will be silenced after you press the
ALARM SILENCE key (unless a new alarm comes in)
Alarm Suspend Time
Time that alarms will be suspended after you select the
Alarm Suspend item in the top level Setup menu
Can Suspend All Alarms
This setting controls if the clinician is able to suspend all
alarms including lethal arrhythmias (Asystole and VFib).
Alarm Validation
Enables/disables alarm validation (if enabled, limit alarms
are not annunciated until values have been out of limits for
a set period of time see Managing Alarms chapter)
Second Speaker Time
Time that elapses while an alarm tone in sounding before
the second speaker starts sounding
Note -- When a communications link with the MPC is established, these settings will be
overridden by those that are received from the MPC.