xiv Nightingale PPM3 User's Guide April 2019
WARNING ARRHYTHMIA PATIENTS. The Nightingale PPM3 is
designed to operate in the presence of cardiac arrhythmias.
However, the heart rate meter may be adversely affected in some
cases.
WARNING DEVICE INTERCONNECTIONS. Through its OMD (Other
Medical Device) port, the Nightingale PPM3 can be connected to
external devices. The following precautions should be taken:
The OMD cable should not be applied to the patient.
Connected devices should be located outside of the patient
vicinity (greater than 1.5 meters) if they do not comply with IEC
60601-1.
The PPM3 should not be connected to devices that are not
described in this manual.
The over-all system leakage current should be tested and
should comply with IEC 60601-1-1.
Nightingale MPC to PPM3 Network installation must be
performed by service personnel that are authorized by Zoe
Medical. Connection of the PPM3 to a network that includes
other equipment could result in previously unidentified risks.
WARNING BATTERY HANDLING. The Nightingale PPM3 contains a
lithium ion coin cell battery and transport battery pack. The following
precautions should be taken regarding these batteries:
Do not immerse in water.
Do not heat or throw in fire.
Do not leave in conditions over 60 ºC or in a heated car.
Do not attempt to crush or drop.
Only use the battery pack with the Nightingale PPM3.
Follow the instructions in the Disposal chapter of this manual
when the PPM3 is taken out of service.
WARNING AUDIBLE TONES. The Nightingale PPM3 should sound
audible startup tones whenever it is powered on (two tones followed
by two higher beeps). If a unit does not sound the startup tones
when it is powered on, remove the unit from service and contact Zoe
Medical Technical Support.