WARNING – Patient monitoring will be suspended at both the PPM3
and MPC for this patient when the Enter Standby Mode menu item is
selected.
Select the "Transport Patient" menu item if you are about to disconnect the PPM3
from the MPC in order to move the patient. This will place the MPC in Transport
Mode for this patient and will prevent Connection Lost alarms from being
sounded at both the PPM3 and MPC.
WARNING – Patient monitoring will be suspended at the MPC for this
patient when the Transport Patient menu item is selected.
When the PPM3 is reconnected to the MPC, the MPC will automatically exit
Transport Mode and resume monitoring.
Select the "Discharge Patient" menu item if you are finished monitoring this
patient. This will de-assign the patient from both the PPM3 and MPC (although
the patient’s trend data will still be archived at the MPC). In addition, the PPM3
will restore the monitor’s parameter and waveform settings to the Department
Defaults (last saved setup), and suspend monitoring (as described above).
WARNING – Patient monitoring will be suspended at the MPC for this
patient when the Discharge Patient menu item is selected.