22-6 Nightingale PPM3 User's Guide April 2019
Maintenance and Storage
PPM3 Function
Procedure
IBP Invasive Blood Pressure
The IBP (P1 and P2) interfaces do not require calibration during normal
clinical operation. In order to calibrate the IBP interfaces, perform the
following:
1. Connect a static pressure source, such as ambient air, to pressure
transducer that is connected to the P1 invasive pressure interface.
2. Select "Start P1 Calibration," after first pressing the SETUP key, then
"Administration," then "Service,” then "Calibration” (password required to
access this screen).
3. Await "P1 Calibration OK" message, or similar if the interface is labeled
something other than P1.
4. Repeat for interface P2.
NBP Non-Invasive Blood
Pressure
The NIBP interface does not require calibration during normal clinical
operation. The accuracy of the NIBP pressure sensor and overall functionality
of the NIBP interface can be verified as detailed in the Nightingale PPM3
Functional Tests. If further calibration is required, contact Zoe Medical
Technical Support.
If there is a failure in one of the above procedures, please contact Zoe Medical
Technical Support. If a PPM3 unit needs to be returned to the factory for repair,
Technical Support will provide a return authorization number.
Email: customersupport@zoemedical.com
Phone : (978) 887-4013