April 2019 Nightingale PPM3 User's Guide 24-1
Technical Data
24. Technical Data
General
Dimensions
11.3”W x 7.2”H x 2.4”D (288 mm x 182 mm x 60 mm)
Weight
4.5 lb (2.0 g), 5.0 lb w/ Oridion Microstream® CO
2
(2.3 kg)
Finish
PC/ABS
Power Requirements
100 240 VAC, 1.2 A max
Mains Frequency Range
50 60 Hz
Power Consumption
12W nominal, 30W (when charging battery)
Standards Conformance
IEC 60601-1:2005 (General Safety)
AAMI ES60601-1:2005 (General Safety)
CSA C22.2#60601-1:2008 (General Safety)
IEC 60601-1-2:2007 (Class B) (EMC)
IEC 60601-1-4:2000 (General Safety)
IEC 60601-1-8:2006 (Alarms)
IEC 60601-2-27:2011 (ECG Monitoring)
AAMI SP10:2002/A1:2003 (Non-Invasive Blood Pressure)
ISO 80601-2-30:2009 (Non-Invasive Blood Pressure)
IEC 60601-2-34:2011 (Invasive Blood Pressure)
ISO 80601-2-55:2011 (CO
2
Respiratory Gas Monitoring)
ISO 80601-2-56:2009 (Temperature)
ISO 80601-2-61:2011 (SpO
2
)
IEC 62366:2007 (Usability)
IEC 62304:2006 (Software)
Patient Risk Current
(IEC 60601-1)
Electromedical Apparatus with Isolated Patient Connection.
Meets the following limits:
Enclosure Risk Current < 100 μA
Patient-applied Risk Current < 10 μA
Patient Isolation Risk Current < 50 μA
Earth Risk Current < 500 μA
Type of Protection (Electrical)
Class I
Degree of Protection (Electrical)
Type CF, Defibrillation-proof
Degree of Protection (Water)
Ordinary Equipment (IPX1)
Disinfecting Method
Per the instructions in the Cleaning chapter
Degree of Safety (Flammable
Anesthetic Mixture)
Not suitable for use in the presence of a Flammable Anesthetic
Mixture
Mode of Operation
Continuous