3-4 Nightingale PPM3 User's Guide April 2019
ECG and Heart Rate Monitoring
CAUTION To avoid large offset potentials due to polarization of
electrodes, be sure to use pregelled, Ag/AgCl disposable electrodes, and
be sure that all electrodes used on a patient are of the same type.
Note The HR/PR numeric is displayed in green when sourced from ECG (default). The
color changes to cyan or green respectively when sourced from SpO2 or an ART-labeled
IBP.
Note If an ECG waveform is not displaying on the PPM3, follow the instructions in the
chapter Working With Menus to select one or more ECG waveforms for display.
Note Visually inspect the ECG cables on a daily basis and follow the instructions in the
Cleaning and Maintenance chapters as needed. Also, check for an inoperable ECG
circuit which would manifest itself as a blank ECG waveform and a "HR lead off"
message.