7-8 Nightingale PPM3 User's Guide April 2019
Oridion CO
2
Monitoring
Message
Parameter
Value
Possible Cause
Suggested Action
IPI age range not set
---
For pediatric patients,
the age range must be
set to correctly determine
IPI.
Set the age range in the CO
2
setup
screen.
CO
2
Unplugged
---
CO
2
sampling line is not
connected.
Connect the CO
2
sampling line to the
PPM3.
Disable CO
2
monitoring in the PPM3
Parameters menu if these parameters
are no longer clinically required.
CO
2
Occluded Line
---
CO
2
sampling line cannot
be cleared due to
moisture or other
obstruction.
Replace the sampling line. If connected
to scavenging system, disconnect to see
if message disappears.
CO
2
Purging Line
---
Microstream module is
trying to clear excess
humidity
If the message disappears, resume
usage. If the purging operation is
unsuccessful or the Occluded Line
message appears, disconnect the line
and reattach. If the problem continues,
replace the sampling line.
CO
2
problem detected
---
CO
2
interface has
encountered a problem.
Check the CO
2
subsystem including
sampling line and exhaust port. Verify
that the exhaust port is not blocked.