8-12 Nightingale PPM3 User's Guide April 2019
ISA CO
2
sampling line
cannot be cleared due to
moisture or other
obstruction.
Replace the sampling line. If connected
to scavenging system, disconnect to see
if message disappears.
ISA CO
2
sampling line
has become unplugged
from the LEGI connector
on the front of the
module.
Reconnect the sampling line.
1) The lens within IRMA
airway adapter has
become fogged or
blocked.
2) IRMA airway adapter
has become unplugged
from the sensor.
Replace or reconnect the airway adapter.
CO
2
zero offset needs to
be removed.
Follow the zeroing procedure provided in
this chapter.
ISA module or IRMA
sensor have reported an
internal error.
Disconnect the module or adapter from
the PPM3 and then reconnect. Contact
technical support if the problem
continues.
CO
2
interface has
encountered a problem.
Check the CO
2
subsystem including
sampling line and exhaust port. Verify
that the exhaust port is not blocked.
Power-cycle the PPM3. Contact
technical support if the problem
continues.