60 BM series
Outputting the clock data of the external device āinfo 3ā
When the "GLP output (info)" parameter of the function table is set to "3", the clock data of
external devices such as a printer or a personal computer can be used instead of the balance
built-in clock data for the GLP/GMP compliant report. This is useful for standardizing the clock
data using the clock function of an external device.
Notes
To output the clock data of an external device, the external device such as an AD-8127
printer or Windows communication tools software WinCT RsCom must have a function to
output the time and date when <ESC>D or <ESC>T is received.
Even withinfo 3 set, the data memory function saves the balance built-in clock data for
the calibration report.
Manufacture