BM series 61
Calibration test report using the internal mass
Note Calibration test does not perform adjustment.
Key operation
1 Press and hold the CAL key (for approx. 2 seconds) to display CC in and release the key.
2 CC is displayed and the balance is checked automatically.
3 The zero point is measured and the weight value is displayed for a few seconds.
4 The internal mass is weighed and the weight value is displayed for a few seconds.
5 glp is displayed and the calibration test report is output.
6 The balance returns to the weighing mode.
Command
This calibration test can be performed with the TST command.
Function setting "info 1" Function setting "info 2"
AD-8121 format
General format using the balance built-in clock data
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR Carriage return, ASCII 0Dh
LF Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test type
Zero point value
Target weight value
Target weight
Signature
-----------A-&-D
MODEL-----BM-300
S/N-----01234567
ID------LAB-0123
DATE--2011/02/20
TIME----12:34:56
CAL.TEST(INT.)
ACTUAL
-------0.0000--g
----+200.0002--g
TARGET
----+200.0000--g
SIGNATURE
---------------