62 BM series
Calibration report using an external weight
Key operation
1 Press and hold the CAL key (for approx. 2 seconds) to display Cal out and release the key.
2 Cal 0 is displayed.
3 To update the calibration weight value, press the RANGE key and proceed to Step 4.
To use the preset calibration weight value, proceed to Step 5.
4 Specify the calibration weight value using the following keys.
RANGE key
........
To select the blinking digit.
RE-ZERO (+)key
...
To increase the value of the blinking digit.
MODE (-)key
.......
To decrease the value of the blinking digit.
PRINT key
.........
To store the new value.
5 Press the PRINT key. The zero point is measured and the weight value is displayed for a few
seconds.
6 Place the displayed weight on the pan. Press the PRINT key to measure the weight and the
weight value is displayed for a few seconds.
7 end is displayed. Remove the weight from the pan.
8 If GLP output is used, glp is displayed and the calibration report is output.
9 The balance returns to the weighing mode.
Function setting "info 1" Function setting "info 2"
AD-8121 format
General format using the balance built-in clock data
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR Carriage return, ASCII 0Dh
LF Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration weight
Signature
-----------A-&-D
MODEL-----BM-300
S/N-----01234567
ID------LAB-0123
DATE--2011/02/20
TIME----12:34:56
CALIBRATED(EXT.)
CAL.WEIGHT
----+200.0000--g
SIGNATURE
---------------