6243/44 DC Voltage Current Source/Monitor Operation Manual
4.3 Operation in the Pulse Source Mode
4-13
The measurement point changes depending on the specified Measure Delay time.
For details, refer to Section 4.9.1, “(2) Restrictions on Time Parameters.”
The measurement point in the Pulse source mode becomes as follows:
(a) Free-run during Operate status
• The pulse wave is repeated with the cycle of (the specified Period + Source Delay time).
• If the source value is changed, the current pulse operation stops and a pulse wave with a new base
value and source value is generated.
(b) Hold during Operate status
• Source value change is the same as (a).
• Triggers input during a Period and during Source Delay time are ignored.
(c) Change from Operate status to Standby status in the free run mode, and source value changes including
range changes
• When Operate is specified, the 6243/44 waits for the Operate processing time + Source Delay time
and then generates the pulse.
• The source value change including range changes during operation waits for the Range change pro-
cessing time + Source delay time and then the pulse wave is generated.
• For source value changes including range changes, the base value and the pulse value are changed to
be in the same range.
(d) Change from Operate status to Standby status in the hold mode, and source value changes including
range changes
• When a trigger is input during range change processing, the 6243/44 waits for processing comple-
tion and then the pulse is generated.
• Triggers input during Period, Source Delay time, and Operate processing are ignored.