6243/44 DC Voltage Current Source/Monitor Operation Manual
8.1.4 Calibration Points and Tolerance Range
8-4
8.1.4 Calibration Points and Tolerance Range
Calibrate the 6243/44 using measuring instruments satisfying the required accuracy outlined in Section
8.1.1, “Cables and Measuring Instruments Required for Calibration” meeting the tolerance ranges shown
in the following table.
8.1.5 Calibration Procedure
Figure 8-2 shows an overview of the calibration procedure.
Figure 8-2 Overview of the Calibration Procedure
Range
Source calibration Measurement calibration
Calibration point
Tolerance
range
Calibration point
Tolerance
range
Zero Full scale Zero Full scale
320 mV
3.2 V
32 V
110 V
0 V
300 mV
3 V
30 V
100 V
10 V
100 V
1 mV
10 mV
0 V
300 mV
3 V
30 V
100 V
10 V
20 V
200 V
2 mV
32 A
320 A
3.2 mA
32 mA
320 mA
2 (1 A)
0 A
30 A
300 A
3 mA
30 mA
300 mA
1 A
1 nA
10 nA
100 nA
1 A
10 A
100 A
0 A
30 A
300 A
3 mA
30 mA
300 mA
1 A
500 pA
5 nA
50 nA
500 nA
5 A
30 A
VS Calibration
VM Calibration
IS Calibration
IM Calibration
Save CAL-Data
Init to Default
Calibration items
Parameter items
Parameter Group
SWEEP
SYSTEM
MENU
Line Frequency
Cal Switch Set
Calibration
Execute
Execute
Range/Set
Range and polarity
settings
Zero point
calibration
Full scale
calibration
Zero CAL Full CAL
Save?
ON