9.1.5 General Specifications
9-11
Hold time:
Auto range delay time:
9.1.5 General Specifications
Operating environment conditions: Ambient temperature 0ï‚°C to 50ï‚°C.
Relative humidity: 85% or lower (with no condensation)
Storage environment conditions: Ambient temperature -25ï‚°C to +70ï‚°C.
Relative humidity: 85% or lower (with no condensation)
Warm-up: 60 minutes or longer (until the specified accuracy is reached)
Display: 5 ï‚´ 7 dot matrix fluorescent character display tube
Power supply: AC power, 100 V, 120 V, 220 V, 240 V (user switchable)
Line frequency: 50 Hz/60 Hz
Power consumption: 340 VA or less
External dimensions: Approximately 212 (W) ï‚´ 177 (H) ï‚´ 450 (D) mm
Mass: 15 kg or less
Setting range Resolution Accuracy
3 ms to 60000 ms 1 ms 2% + 1 ms)
Setting range Resolution Accuracy
0 ms to 500 ms 1 ms 5% + 1 ms)
Option No. Standard opt. 32 opt. 42 opt. 44
Power supply voltage 100 V 120 V 220 V 240 V