6243/44 DC Voltage Current Source/Monitor Operation Manual
4.5.3 Source Range
4-24
Set values and decided ranges for auto range
Taking the example of DC sweep, Table 4-9 shows how the source range and source data become re-
sults of the range specification for the same source value.
Setting value
Specified
range
6243
Voltage
source
0 V |V| 320 mV
320 mV |V| 3.2 V
3.2 V |V| 32 V
32 V |V| 110 V
320 mV
3.2 V
32 V
110 V
Current
source
0 A | I | 32 A
32 A | I | 320 A
320 A | I | 3.2 mA
3.2 mA | I | 32 mA
32 mA | I | 320 mA
320 mA | I | 2 A
32 A
320 A
3.2 mA
32 mA
320 mA
2 A
6244
Voltage
source
0 V |V| 320 mV
320 mV |V| 3.2 V
3.2 V |V| 20 V
320 mV
3.2 V
20 V
Current
source
0 A | I
| 320 A
320 A | I | 3.2 mA
3.2 mA | I | 32 mA
32 mA | I | 320 mA
320 mA | I | 3.2 A
3.2 A | I | 10 A
320 A
3.2 mA
32 mA
320 mA
3.2 A
10 A
Table 4-9 Range and Data Differences in Accordance With Range Specification
(Example: Showing DC Sweep With 6243)
Source setting
Fixed range
(110 V range fixed)
Auto range
Parameter
Setting
value
RangeDataRangeData
Bias value
Start value
Stop value
0.1 V
2 V
80 V
110 V
110 V
110 V
0.10 V
2.00 V
80.00 V
320 mV
3.2 V
110 V
100.00 mV
2.0000 V
80.00 V
• 110 V range is selected for all.
• 320 mV range is selected for Bias values.
3.2 V range is selected for Start values.
110 V range is selected for Stop values.
320 mV range to 110 V range is used during sweep.