6243/44 DC Voltage Current Source/Monitor Operation Manual
9.2.3 Source and Measurement Functions
9-20
9.2.3 Source and Measurement Functions
DC source/measurement : Source/measurement of DC voltage/current
Pulse source/measurement : Source/measurement of pulse voltage/current
Pulse LO value gets the same polarity as HI value.
DC sweep source/measurement : Source/measurement by Linear, log, and random
Pulse sweep source/measurement : Source/measurement by Linear, log, and random
Pulse LO value gets the same polarity as HI value.
Sweep mode : Reverse ON (Round-trip) / OFF (One-way)
No. of sweep repeats : 1 to 1000 times, infinite
Max. number of sweep steps : 5000 steps
Max. memory for random sweep : 5000 data items
Measurement data buffer memory : 5000 data items
Calculation functions : NULL calculation
Comparison calculation (HI/GO/LO)
Trigger method : Automatic trigger (DC free run/Pulse repeat)
Source/measurement by external trigger
Output terminals : Front; safety socket
HI OUTPUT, HI SENSE, LO OUTPUT, LO SENSE
Max. source voltage between terminals : (HI - LO): 20 V peak MAX
(OUTPUT - SENSE): 2 V peak MAX
(LO - Case): 250 V peak MAX
Max. remote sensing voltage :
;
1 V MAX
HI OUTPUT - HI SENSE, LO OUTPUT - LO SENSE
(the voltage between HI SENSE - LO SENSE must be
within the range of the maximum output voltage)
Voltage measurement input resistance : 1 G or higher
Voltage measurement input leakage current : 2 nA or lower
GPIB : Complies with IEEE-Std. 488-1978
Interface functions SH1, AH1, T5, L4, SR1,
RL1, PP0, DC1, DT1, C0, E2
Single-wire signal : TRIGGER IN
SYNC OUT
COMPLETE OUT/BUSY IN/BUSY OUT
INTERLOCK/OPERATE IN/OPERATE OUT