6243/44 DC Voltage Current Source/Monitor Operation Manual
4.9.2 Measure Delay and Settling Time
4-45
•6244
• When the current limiter range is 320 A to 10 A: Ts = 0.3 + (0.8 + 7500/DIL) Vs/100 [ms]
Example: Ts unit (ms)
(2) Current source
The 6243/44’s settling time (Ts) becomes, in accordance with the current source change value (Is),
the current sense resistance (Rs) and the load voltage, (V
RL
= R
L
· Is).
Set the Measure Delay (Td) higher than Ts.
•6243
Example 1: When 0.2 mA is applied to a 100 k resistor in the 3.2 mA range
Is = 0.2 mA
V
RL
= 0.2 mA 100 k = 20 V
28 V
RL
= 28 20 = 560 s
9 V
RL
/ (Rs· Is) + 200 = 9 20 / (220 0.2 mA) + 200
= 4291 s
From this follows that Ts = 4291 s
Make setting so that Td 4.29 ms
Current limiter
range
Vs [V]
Current limiter setting DIL(digits)
300 3000 5000 10000 32000
320 A to 10 A
1 0.6 0.3 0.3 0.3 0.3
10 2.9 0.6 0.5 0.5 0.4
20 5.5 1.0 0.8 0.6 0.5
Ts value Unit [s]
Range
|V
RL
| 10 V
|V
RL
| 10 V
|Is| 18000 digits |Is| 18000 digits
32 A35 V
RL
/ (Rs· Is)
320 A to 2 A 300
Either 28 V
RL
or 9 V
RL
/ (RsIs) + 200 which is
greater the other.
Rs value
Range Rs []
32 A
320 A
3.2 mA
32 mA
320 mA
2 A
22 k
2.2 k
220
22
2.2
0.22