6243/44 DC Voltage Current Source/Monitor Operation Manual
5.2.11 Program Examples
5-37
Call SUBsend(vig, “H”) Output OFF
Measurement buffer memory data recall
Call SUBsend(vig, “RN1,0”) Setting to measurement buffer memory recall mode and specifying
recall address from 0
Do Infinite loop
Call SUBread(vig, dt) Measurement buffer memory data recall
Outputting memory data by data recall after memory recall mode
setting, adding recall number by 1
If 1 = InStr(1, dt, “EE +888.888E+8”) Then
Exit Do If recalled data is empty, exiting infinite loop
End If Ending If
Loop Ending Do
Call SUBsend(vig, “RN0,0”) Releasing measurement buffer memory recall mode
Call ibonl(vig, 0) Setting device (6243/44) to offline
End Sub Ending event procedure
Subroutine
Private Sub SUBread(vig As Integer, dt As String) Recalling talker data
Call ibrd(vig, dt) Recalling talker data
Text1.SelStart = Len(Text1.Text) + 1 Specifying display position for text box (Text1)
Text1.SelText = dt & vbCrLf Displaying measurement data in text box (Text1)
End Sub
Subroutine
Private Sub SUBsend(vig As Integer, cmd As String) Sending command character string
Call ibwrt(vig, cmd & Chr(10)) Sending command character string + terminator LF(Chr$(10))
End Sub