6243/44 DC Voltage Current Source/Monitor Operation Manual
9.1.1 Voltage/Current Source
9-5
Settling time: The time to reach, from the start time, 0.03% of the final value, by changing the output
from zero to full scale. However, in the case of a purity load resistor, load capacity will
be below 2.5 pF, source value, and the limiter setting will be at full scale.
Line regulation: 0.003% of range
Load regulation: Voltage source: Within 0.003% of range at the maximum load with 4 Wire con-
nection
Current source: Depends on CMV section (A Vo/1 V) of overall accuracy.
Output resistance: 2 Wire connection excluding output cables
Maximum load capacity:
Maximum load capacity at which oscillation does not occur during voltage source or
when the voltage limiter is operating.
Range Settling time
Voltage source 320 mV
300 or less
3.2 V
32 V 700 or less
110 V 2 m or less
Current source 32 A5 ms or less
320 A
3 ms or less
3.2mA
32 mA
320 mA
2 A
Range
Output resistance
Maximum load
capacity
Voltage source Current source
32 A
320 A
3.2 mA
32 mA
320 mA
2 A
500 m or lower
100 m or lower
10 m or lower
10 m or lower
10 m or lower
10 m or lower
1 10
9
or higher
1 10
9
or higher
1 10
8
or higher
1 10
7
or higher
1 10
6
or higher
1 10
5
or higher
1 F
1 F
100 F
100 F
2000 F
2000 F
Resistance of standard accessory cable: 100 m or lower