6243/44 DC Voltage Current Source/Monitor Operation Manual
2.2.5 Sweep Mode Measurement
2-29
Turning OFF the Memory Store
36. Press the MENU key. Select MEASURE BUFFER in the parameter group by
rotating the Data knob. Set the Memory Store parameter to OFF. (SLOT in the
indicator section goes out.)
NOTE: When the memory is not used, set Memory Store to OFF.
Clearing the memory
37. Select MEASURE BUFFER in the parameter group and select the Memory
Clear parameter.
38. Press the key.
39. Press the DIRECT key.
40. Press the EXIT key.
The measurement screen returns.
AZERO
Memory Clear
Execute (Press DIRECT Key)
AZERO
Memory Clear
Execute Done