1569‐PT060EE1 Rev.7 Page47of62
5.5.8 Full scale adjustment D/A
Tocontinue,connectadigitalmultimetertotheterminalsoftheinstrument'sanalogue
output,payingattentiontomaketheproperconnections.
Set‐upthemultimetertomeasurevoltageorcurrent,accordingtothetypeofoutputofT060E.
Toaccessthecalibrationmenu,youmustentertheinstrument’smanagementmenubyselectingthefollowing
function:
C
L
PressKey
O F F S 1
Presskeyuntilvisualizingtheparameter:
A N
S P
Bypressingkey ,theinstrumentbeginsthezeroadjustmentprocedureoftheanalogueoutput.
A N
S P
0
Anumericvalue,proportionaltotheoutputvalueundervoltageorcurrentreadonthemultimeter,isvisualizedinthis
phase.
Presskeystomodifythisvalueuntil10Vor20mAareexactlyreadonthe
multimeter.
Presstoconfirmthedisplayedvalueandpresstocancel.