1569‐PT060EE1 Rev.7 Page8of62
2.2 Transmitter’s functions
InstrumentT060Eisaweighttransmitterwithpotentialthresholdrelaymanagementfunction.
Therefore, it is not a traditional weighing instrument, since Gross, Net and Tare management functions are not
available.
Only ZERO corrections can be executed (through external inputs) with the purpose to recover zero derivatives or
deviationsrelativetotheweighingsystem.
The onboard relays are managed as alarm relays, therefore they
are always kept energized and are de‐energized in
caseofalarm.
2.2.1 Management of the calibration
T060Eforeseestheset‐upofatheoreticalcalibration.However,theoperatorcanexecutepotentialcorrectionsduring
thecalibrationphase.Theadvantageofthistypeofset‐upliesinthefactthatthesystemcarriesoutanautomatic
calibration,byknowingthecells’characteristics.
Thesystemisnot
validincaseofloadcellsnotsupportingdirectlytheentireweightappliedtothesystem(e.g.incase
ofaweighingpinappliedtoaresultantofareturnsystem).Inthiscase,theoperatormustsettheweightvalueused
attheendofthemeasurementchain.
Forexample,
incaseofaweighingpinappliedtothefixedcomponentofaliftingsystem,theoperatormustsetthe
valueactuallyliftedbythebridgecrane;forexample,incaseofpinweighing1tonandthecapacityofthebridge
craneconsistingof300tons.Inthiscase,
thesignalwillbesetaccordingtothepin’sdatasheetandthecapacitywill
besetto300(correspondingto300tonsofthesystem'scapacity),whilethesensitivitywillbesetaccordingtothe
accuraciesidentifiedintheapplicationofthesystem(forexample,ifitweights200kg,
thesensitivitymustbesetto
0.2).
Theresolution(divisionvalue)issettotheresultof the60000divisions.Higherset‐upsarediscardedbythesystem.
Thesystem’sresolutionisgivenbytheratiobetweencapacityandresolution.
Asystemwitha100Kgcellandaresolution
of0.01isasystemconsistingof10000divisions.
S E G N 1
T060Eforeseesloadcellswithmaximumsignalof3.5mV/V.
P O R T 1
The value set in this parameter constitutes the full scale of the weighing system. It foresees a maximum value of
100000divisions.
S E N S 1
Thevaluecanbeselectedwithintherangefrom50to0.001divisions.
The change of both system capacity and load cell signal (mV/V) cause the clearance of calibration data
(linearizationsincluded)alreadycarriedout,ofsetpointsvaluesandtheanalogoutputdata.