Configuration and Activation 4-63
Qstart
XML
configuration screen. Choose Depth to initiate sampling
based on UNIDEPTH readings from the selected depth combo
sensor. Flow Rate and Flow Total are instantaneous calculations
based on the selected depth and velocity sensors in the Combo
Sensor to Use.
z Trigger Enter the sampling trigger threshold corresponding to the
selected Type for which you want the sampler activities to
initiate. For example, if Depth is the selected Type, enter the
corresponding threshold value in inches (or millimeters) at which
you want the monitor to pulse the sampler.
z Samples per wakeup Select and enter or use the up/down arrows
to indicate the total, consecutive number of times you want the
monitor to pulse the sampler within a single interval each time the
trigger condition is met.
z Pulse Length Enter the duration of the pulse you want to send to
the sampler once sampling activities begin.