z Store Data Select this checkbox to ensure the monitor logs
pressure depth data to memory. This option is selected by
default.
z Automatic Calibration (only applicable to ParaFlow
sensors) Select this checkbox to ensure the monitor
automatically calibrates questionable or erroneous pressure
depth data to read consistently the Pipe Height minus the
Physical Offset when the manhole is not surcharged.
Note: Do not reactivate the ADS TRITON+ monitor
when the manhole is surcharged. The Pressure sensor will
not work properly if this is done. If it is imperative that the
monitor be reactivated, disable Automatic Calibration
before reactivating.
Temperature tab – ParaFlow/Surface Combo
Edit the parameters corresponding to the temperature sensor in the
ParaFlow or Surface Combo Sensor in the same way in which you
edited the parameters for the temperature sensor in the Peak Combo
Sensor. Refer to the parameter descriptions and instructions for the
Temperature tab on page 4-60 for detailed information on the
individual parameters on the Temperature tab.