11-8 ADS TRITON+ Manual
Codes Initiate after Removing Magnet from Contact with Monitor Top
3. Upon activation, interpret the diagnostic codes in succession in
the following way:
Note: Certain diagnostic information may not be relevant
or available based on the type of communication used by
the monitor. When no information is available for a
particular communication type, the applicable codes will
display zeros.
z indicates modem initialization
z indicates code initiation
z is followed by the modem’s model number
z is followed by the wireless provider’s number
z is followed by the battery voltage to two decimal
places