Stop positions
The system includes stop positions for manual movement of the X-ray tube
head.
The preferred positions of the stops are defined during installation.
The stop positions are used to position the system manually for typical
radiographic examinations, e.g. a SID of 180 cm for chest examinations.
Stop positions are different for radiographic table and radiographic wall
stand. Which stop positions are active depends on the orientation of the X-ray
tube.
To enter a stop position, move the X-ray tube head unit using the control
buttons. The movement is stopped when a stop position is reached. Move at a
regular speed to prevent the X-ray tube head unit to skip the stop position.
To leave a stop position, release and press again the according movement
control button.
130 | Valory (system with floor mounted column) | Positioning the X-Ray Tube