X-Ray Tube Technical Data
Manufacturer Canon Electron Tubes & Devices Co., Ltd.
1385 Shimoishigami
Otawara-Shi, Tochigi-Ken 324-8550 Japan
E7884X X-ray Tube 12º
150 kVp
dual focal spots 0.6 and 1.2 mm
300 KHU
LS 20/50 kW (50Hz) 22/54 kW (60Hz)
7,24x10
6
mAh@150kVp maximum load
E7252X X-ray Tube 12º
150 kVp
dual focal spots 0.6 and 1.2 mm
300 KHU
LS 14/41 kW (50Hz) 16/45 kW (60Hz)
HS 27/75 kW (180Hz)
7,24x10
6
mAh@150kVp maximum load
E7254FX X-ray Tube 12º
150 kVp
dual focal spots 0.6 and 1.2 mm
400 KHU
LS 22/55 kW (50Hz) 23/60 kW (60Hz)
HS 40/102 kW (180Hz)
9,66x10
6
mAh@150kVp maximum load
Exposure parameters
Tube Voltage
The tube voltage can be selected in steps of 1 kV in the range of 40 to 150 kV.