Safety directions for the floor mounted column
If the radiographic table is installed in the movement range of the floor
mounted column, make sure that the X-ray tube, collimator or X-ray tube arm
do not collide with the tabletop, especially when the X-ray tube moves under
the tabletop.
If the radiographic wall stand is installed in the movement range of the floor
mounted column, make sure that the X-ray tube, collimator or X-ray tube arm
do not collide with the radiographic wall stand.
80 | Valory (system with floor mounted column) | Safety Directions