DR Detector is Exceeding the Maximum Working
Temperature
Details A message is displayed on NX indicating that the DR De-
tector is exceeding the maximum working temperature.
Cause Due to ambient temperature conditions and the number
of acquired images, the DR Detector’s internal tempera-
ture may become too high.
Brief Solution
1. Power off the DR Detector.
2. Leave the DR Detector unpowered for at least one
hour.
3. Stop the NX workstation.
4. Power on the DR Detector.
5. Start the NX workstation.