Dose Area Product Meter (IBA DAP) Technical Data
Manufacturer
IBA Dosimetry GmbH
Bahnhofstrasse 5
DE-90592 Schwarzenbruck
Supported Type 120-131 HS/RS485
Dose area product range (0.1...99999999.99) cGy x cm
2
DAP resolution 0.01 cGy x cm
2
Active area 14.0 cm x 14.0 cm
Dimensions 17.9 cm x 16.6 cm x 1.7 cm (WxDxH)
Weight approx. 220 g
Equivalent filtration of the ionization
chamber at 70 kV
0.31 mm Al
Correction factors for using the DAP meter on high altitude
Environmental conditions Correction factor
75 kPa (ca. 2500 m)
0° Celcius
1.26
75 kPa (ca. 2500 m)
20° Celcius
1.35
70 kPa (ca. 3000 m)
0° Celcius
1.35
70 kPa (ca. 3000 m)
20° Celcius
1.45