The use of the DR detector bucky is restricted to detectors with a size
equivalent to 14x17 inch (35x43 cm) or to 17x17 inch (43x43 cm).
Alternative components:
Radiographic table with fixed height
Radiographic wall stand with fixed DR detector
Limited configurations:
configuration without radiographic wall stand
configuration without radiographic table
configuration with only a floor mounted column with X-ray tube head
X-ray parameters are controlled using the software console on the MUSICA
Acquisition workstation. The software console is available on the MUSICA
Acquisition workstation, to synchronize the X-ray exposure parameters with
the generator.
Valory can be used in combination with a CR digitizer and CR cassettes. The
use of the bucky is restricted to CR cassettes with a size of 14x17 inch (35x43
cm) or 17x17 inch (43x43 cm). After each exposure, remove the CR cassette
from the bucky and insert it in the digitizer. The system does not prevent
double exposure of a CR cassette in the bucky.
Valory can be used in combination with a CR digitizer and a CR full body
cassette holder to perform full leg full spine examinations. Refer to the CR Full
Leg Full Spine User Manual (document 4408, part of the MUSICA Acquisition
workstation User Documentation).