MUSICA Ac-
quisition
workstation
(typical,
without UPS
and extra
monitors)
45 W
High voltage output values
High voltage
power out-
put (at 0.1s)
500mA: 80kV
400mA: 100kV
320mA: 125kV
266mA: 150kV
625mA:
80kVp
500mA:
100kVp
400mA:
125kVp
330mA:
150kVp
800mA:
80kVp
650mA:
100kVp
520mA:
125kVp
430mA:
150kVp
800mA:
80kVp
800mA:
100kVp
640mA:
125kVp
530mA:
150kVp
kV-Range 40-150 kV
mAs-Range 0.5-600 mAs
mA-Range 10-500 mA 10-650 mA 10-800 mA
ms-Range 1-6300 ms
Mechanical data
Dimensions 550 mm x 630 mm x 970 mm (WxDxH)
Weight 129 kg
Duty cycle The Generator duty cycle is continuous, but limits should be
set during installation depending on the capacity of the X-ray
tube.
(*) Split phase AC line with two hot phases, e.g. 240 VAC available as two 120
VAC hot phases
188 | Valory (system with floor mounted column) | Product Information
0426A EN 20220804 1858