108 1290 Infinity Quaternary Pump User Manual
7 Error Information
Pump Error Messages
Drive current too low
Error ID: 29205
The current consumption of the pump drive is too low.
• Parameter: 1 – 2 referring to pump drive
Drive current too high
Error ID: 29236
The current consumption of the pump drive is too high.
• Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1 Drive motor defect.
Please contact your Agilent service
representative.
2 Wrong/missing connection of pump drive to
main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Blockage of system before pressure sensor.
Check for blockage of e.g. outlet valve filter frit,
Multi Purpose Valve, heat exchanger.
2 Drive motor defect.
Please contact your Agilent service
representative.