72 1290 Infinity Quaternary Pump User Manual
4 Using the Pump
Solvent Information
Solvent Information
Introduction
Observe the following recommendations on the use of solvents.
• Follow recommendations for avoiding the growth of algae, see “Algae
Growth in HPLC Systems” on page 54.
• Small particles can permanently block capillaries and valves. Therefore,
always filter solvents through 0.4 µm filters.
• Avoid or minimize the use of solvents that may corrode parts in the
flow path. Consider specifications for the pH range given for different
materials like flow cells, valve materials etc. and recommendations in
subsequent sections.
Materials in Flow Path
Following materials are used in the flow path of this module:
Part Materials
Degasser chamber TFE/PDD copolymer, PFA (internal tubings); PEEK (inlets); FEP
(tubings); ETFE (fittings)
Ultra clean tubings
1
1
Ultra clean tubings are available for the use with high-end MS detectors. They are also compatible
to THF.
PFA (tubings), PEEK (fittings)
Microfluidic structures
2
2
Inlet Weaver, Jet Weaver, Heat Exchanger
SST
MCGV PEEK, FEP, PFA, Al
2
O
3
-based ceramic, ruby, sapphire, SST
Passive inlet valve SST, gold, ruby, ZrO
2
-based ceramic, tantalum
Outlet valve SST, gold, ruby, ZrO
2
-based ceramic, tantalum
Pump head SST
Pistons ZrO
2
-based ceramic
Piston/wash seals UHMW-PE, SST
Pressure sensor SST
Multi Purpose Valve Polyimide, SST, DLC