164 1290 Infinity Quaternary Pump User Manual
9 Maintenance
Disassembling the Secondary Pump Head
Disassembling the Secondary Pump Head
CAUTION
Damage of pump piston
The pump piston is made of ZrO
2
-based ceramic, which is a very hard and resistant
material, but it is sensitive to shearing forces from the side.
Do not try to remove the pump piston from the rear.
Do not use the piston for removing pump seals.
1 Remove the outlet filter at the top of the pump head. 2 For disassembling the pump head, remove the 4 hex
screws at the rear of the pump head.