1290 Infinity Quaternary Pump User Manual 113
Error Information
7
Pump Error Messages
Writing the multi purpose valve tag failed
Error ID: 29241
Writing the multi purpose valve tag failed.
Parameter: None
Unknown multi purpose valve type
Error ID: 29242
The type information of the multi purpose valve is invalid.
Parameter: None
Pump drive encoder error
Error ID: 29211
The pump drive encoder has generated no signal.
Parameter: 1 – 2 referring to pump drive
Probable cause Suggested actions
1 Multi purpose valve head tag defect.
Replace multi purpose valve head.
2 Multi purpose valve tag head reader is
defect.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Wrong valve head installed.
Check or replace multi purpose valve head.
2 Valve head has invalid RFID tag content.
Check or replace multi purpose valve head.
Probable cause Suggested actions
1 Pump drive encoder is defect.
Please contact your Agilent service
representative.