196 1290 Infinity Quaternary Pump User Manual
9 Maintenance
Preparing the Pump Module for Transport
Preparing the Pump Module for Transport
1 Flush all solvent channels with isopropanol.
2 Remove solvent inlet tubes from solvent reservoirs and tubing clips at
other modules.
3 If applicable, remove tubings between the seal wash function and
solvent bottle/waste.
4 Remove cable and capillary connections to other modules.
5 Remove the module from the stack.
6 Remove the waste tube.
When If the module shall be transported or shipped.
Parts required p/n Description
9301-0411 Syringe; Plastic
9301-1337 Syringe adapter
G4204-44000 Transport protection foam
CAUTION
Mechanical damage
For shipping the module, insert the Protective Foam to protected the module from
mechanical damage.
Be careful not to damage tubing or capillary connections while inserting the module
in the Protective Foam.

Table of Contents

Other manuals for Agilent Technologies 1290

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 and is the answer not in the manual?

Agilent Technologies 1290 Specifications

General IconGeneral
BrandAgilent Technologies
Model1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals