116 1290 Infinity Quaternary Pump User Manual
7 Error Information
Pump Error Messages
Timeout: Wait for run volume
Error ID: 29181
A target condition (run volume, which is the volume delivered since the
method run start) has been sent to the instrument which should have
been reached within an expected time frame but didn’t. Either the limit,
time frame or the current value of the variable has been modified later
directly or indirectly (for example the flow rate).
Timeout: Wait for Volume
Error ID: 29182
A target condition (volume, which is the delivered flow since the limit has
been set) has been sent to the instrument which should have been reached
within an expected time frame but didn’t. Either the limit, time frame or
the current value of the variable has been modified later directly or
indirectly (for example the flow rate).
Probable cause Suggested actions
1 Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.
Probable cause Suggested actions
1 Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.

Table of Contents

Other manuals for Agilent Technologies 1290

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 and is the answer not in the manual?

Agilent Technologies 1290 Specifications

General IconGeneral
BrandAgilent Technologies
Model1290
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals